JCB-6KR Fully Automatic Karl Fischer Moisture Analyzer (Volumetric Method)
Category:
The JCB-6KR Karl Fischer moisture analyzer is tailored to the operational preferences of Chinese users and features high measurement accuracy, excellent repeatability, and stable performance. It is widely applicable across industries such as chemicals, pharmaceuticals, food, batteries, and plastics, and is also well suited for moisture‑content analysis in universities, quality‑inspection agencies, and research institutions. The instrument delivers rapid and precise measurements of water content ranging from 10 ppm to 100%.
Contact :
Product Details
The JCB‑6KR Karl Fischer moisture analyzer is tailored to the operating habits of Chinese users and features high measurement accuracy, excellent repeatability, and stable performance. It is widely applicable across industries such as chemicals, pharmaceuticals, food, batteries, and plastics, and is also well suited for a variety of moisture‑content applications in universities, quality‑inspection agencies, and research institutions. The instrument enables rapid and precise determination of moisture levels ranging from 10 ppm to 100%.
Key Features
. High-precision titration system
A high-precision titration system driven by a precision lead screw, equipped with an imported stepper motor and featuring a drive resolution of 1/3600 mm, is further paired with a custom‑designed burette. This configuration achieves a minimum titration volume accuracy of 0.25 μL, ensuring the system’s accuracy right at the source.
. A more reliable three-way valve
It employs a genuine Japanese‑imported solenoid three‑way valve that is leak‑proof and features a faster response time due to its absence of mechanical components. Liquid‑path switching takes only tens of milliseconds, and compared with mechanical three‑way valves, it can withstand higher pressures and prevents leakage even in the event of pipeline blockage.
. Convenient and quick operation method
The system provides a comprehensive overview of all menus, with multiple quick-access shortcuts for rapid feature navigation. Its simple, intuitive menu design ensures that even without training, you can become an expert user.
. Intuitive and comprehensive display interface
It displays the current‑cell current in real time, enabling a more intuitive observation of the entire reaction process and more accurate sample measurements. During titration, it provides real-time readings of Karl Fischer reagent consumption, water mass, and moisture content.
. Different methods can be configured based on the sample, making testing more convenient.
Five setting options are available, allowing for tailored configurations—such as titration speed and titration delay—to suit different samples, enabling faster and more accurate measurements.
. Real-time drift detection to reduce environmental impact
A real-time drift detector can continuously monitor drift in response to environmental climate changes and, as needed, allow users to specify whether drift correction should be applied in the analytical method.
. Results are saved automatically, and printing can be done without connecting to a computer.
It can store more than 40 titration results, with a result format fully compliant with GLP standards, and can be directly connected to a printer for record printing.
. Fully enclosed system with automated reagent management, ensuring safety
The fully enclosed operating environment ensures that operators remain in contact with reagents at no point during the procedure, effectively safeguarding their safety. With automated solvent handling and waste‑liquid suction and discharge functions, the operation is simplified, eliminating the need for personnel to handle hazardous reagents.
. Automatic cleaning for greater convenience
One‑click cleaning: The instrument features one‑touch cleaning functions, including power‑on and power‑off rinsing, automatic mixing of Karl Fischer reagent, and automated pipeline flushing, effectively preventing issues caused by the crystallization of Karl Fischer reagent.
Technical Specifications
|
Name |
Technical Specifications |
|
Model number |
S300 |
|
Measurement range |
10 ppm–100% water |
|
Resolution |
1ppm |
|
Polarization voltage |
±2000 mV, effective accuracy: ±1 mV |
|
Polarization current |
0–200 μA Effective accuracy: ±0.1 μA |
|
Minimum liquid inflow |
1 μL (20 mL volumetric tube); |
|
Burette |
20 ml; 10 ml; 5 ml |
|
Burette accuracy |
1/20000 |
|
Repeatability |
≥99.7% (10 mg water) |
|
Maximum error |
<1% |
|
Featured Features |
Customizable endpoint delay and titration delay; one-click cleaning; real-time current display; instrument calibration; overflow protection for waste‑liquid containers. |
|
Display screen |
4.3-inch extra-large LCD display |
|
Operating environment |
Temperature: +5 to +40°C; Humidity: <80% |
|
Power supply |
110-220V 50Hz |
|
Instrument dimensions |
495x330x325mm |
|
Instrument weight |
4kg |
Applicable Standards
GB 6283-1982 “Determination of Water Content in Chemical Products—Karl Fischer Method (General Method)”
GB/T 11133-1989 “Method for Determination of Water Content in Liquid Petroleum Products (Karl Fischer Method)”
GB/T 2441.3–2010 “Methods for the Determination of Urea” — Part 3: Water Content — Karl Fischer Titration Method
GB/T 7380-1995 “Determination of Water Content in Surfactants (Karl Fischer Method)”
GB 10670-1989 “Determination of Trace Water in Industrial Fluorinated Methanes by the Karl Fischer Method”
GB 10670-1989 “Determination of Trace Moisture in Industrial Fluoromethanes”
GB/T 606-2003 “General Method for Determination of Water Content in Chemical Reagents: Karl Fischer Titration”
GB/T 8350-2001 “Denatured Fuel Ethanol”
GB/T 3776.1-1983 “Method for Determination of Moisture in Pesticide Emulsifiers”
Production workshop
Application Cases
Certificate of Honor
Related products
Leave a message
Welcome to leave an online message for inquiries and please keep your contact information up to date. We will get in touch with you as soon as possible.
No. 2, Building A27, Jiahai Zhejiang Merchants Science and Technology Park, No. 1 Shenzhen Avenue, Lianhe Subdistrict, Jingzhou Economic and Technological Development Zone, Hubei Province