JCB33400-02 Diesel Total Contaminant Content Analyzer
Category:
The JCB33400-02 diesel total contaminant content analyzer is designed and manufactured by our company in accordance with the latest standard GB/T 33400-2016, “Method for Determination of Total Contaminant Content in Middle Distillates, Diesel Fuels, and Fatty Acid Methyl Esters.” It is used to determine the total contaminant content in middle distillates, diesel fuels with a fatty acid methyl ester volume fraction not exceeding 30%, and pure fatty acid methyl ester (BD100), as specified by the standard. It is also suitable for samples with a kinematic viscosity at 20°C not exceeding 8 mm²/s or at 40°C not exceeding 5 mm²/s, as well as for samples with a total contaminant content ranging from 12 mg/kg to 30 mg/kg; it can further accommodate samples whose total contaminant content falls outside this range. This instrument is the preferred choice for refineries and testing laboratories seeking to measure the total contaminant content in diesel fuel.
Contact :
Product Details
I. Introduction
Dear customer, thank you sincerely for choosing the products of Jingzhou Chuangbo Instrument Manufacturing Co., Ltd. We are committed to providing you with the best possible service.
II. Overview
The JCB33400-02 diesel total contaminant content analyzer is designed and manufactured by our company in accordance with the latest standard GB/T 33400-2016, “Method for Determining Total Contaminant Content in Middle-Distillate Fuels, Diesel Fuels, and Fatty Acid Methyl Esters.” It is used to determine the total contaminant content in middle‑distillate fuels, diesel fuels with a fatty acid methyl ester volume fraction not exceeding 30%, and pure fatty acid methyl ester (BD100), as specified by the standard. It is also suitable for samples with a kinematic viscosity of no more than 8 mm²/s at 20°C or no more than 5 mm²/s at 40°C, as well as for samples with a total contaminant content ranging from 12 mg/kg to 30 mg/kg; it can further accommodate samples whose total contaminant levels fall outside this range. This instrument is the preferred choice for refineries and testing laboratories seeking to measure the total contaminant content in diesel fuel.
This instrument features a long service life, low noise, minimal vibration, compact size, low cost, easy maintenance, an attractive appearance, and simple operation.
III. Technical Parameters:
1. Power supply: AC 220 V ± 10%, 50 Hz
2. Vacuum level: absolute pressure 2 kPa to 5 kPa
3. Vacuum regulation method: adjustment via a stainless steel needle valve
4. Filter membrane diameter: 47 mm; pore size: 0.7 μm
5. Timer: capable of measuring 30 min ± 1 min
6. Operating environment: relative humidity ≤ 80%, temperature 0–45°C
IV. Functions of Main Components
1. Power: Turn the instrument’s power on or off.
2. Vacuum pressure regulator: Used to adjust the vacuum pump’s pressure, ensuring that the absolute pressure within the system remains within the range of 2–5 kPa (corresponding to an instrument vacuum gauge reading of approximately −0.96 to −0.99 bar).
3. Counter: Hours, minutes, and seconds can each be set from 0 to 99.99. The counter features two rows of small buttons: the two front buttons correspond to the tens and units digits (00–99); the middle button selects H for hours, M for minutes, or S for seconds; and the two rear buttons set the units digit (00–99). For example, setting 30 M 20 means 30 minutes and 20 seconds.
V. Instrument Connection
1. Arrange the sand-core filtration apparatus and the suction flask according to the diagram, and install the filter membrane. Connect the suction flask to the filtration apparatus using a rubber tubing.
2. Connect the grounding terminal on the left side of the instrument to the blue clamp using a conductor to prevent static electricity! (See Figure 1 in standard GB/T 33400-2016.)
3. On the right side of the instrument, connect a copper wire to the grounding terminal; extend it into the filtration flask and route it through the gas‑line tubing throughout the apparatus to prevent the buildup of static electricity during the experiment. (See Figure 1 in standard GB/T 33400‑2016.)
4. Connect the vacuum pump to the vacuum pump port on the lower right side of the instrument using a rubber hose. Attach the suction port on the upper right side of the instrument to the side opening of the suction bottle with another rubber hose.
VI. Instrument Operation
Before use, operators must be familiar with the GB/T 33400-2016 standard “Determination of Total Contaminant Content in Middle-Distillate Fuels, Diesel Fuel, and Fatty Acid Methyl Esters,” as well as the safety precautions specified therein.
1. First, set the timer and adjust it to 30 minutes (30M00).
2. Turn on the vacuum pump’s power switch to start the pump.
3. Turn on the instrument’s power switch (the power indicator light will illuminate) and start the instrument.
4. Adjust the vacuum pressure regulator on the instrument to the required pressure range (-0.96 to -0.99 bar).
5. According to the standard, transfer the prepared sample from the filtration funnel into the apparatus for filtration, and begin the experiment.
VII. Issues and Solutions
|
Fault phenomenon |
Solution Method |
|
The vacuum pump fails to start. |
Vacuum pump malfunction: inspect and repair. |
|
Absolute pressure is insufficient. |
1. Check all pipe joints for air leaks. |
|
The instrument is not powered on. |
1. Fuse blown; inspect and replace. |