JCB5096-02 Type Copper Strip Corrosion Tester
Category:
The copper strip corrosion tester is designed and manufactured in accordance with the People’s Republic of China petrochemical standard GB/T 5096, “Test Method for Copper Strip Corrosion of Petroleum Products,” and is suitable for determining the degree of copper corrosion caused by aviation gasoline, jet fuel, motor gasoline, natural gasoline, or other hydrocarbons with a Reid vapor pressure not exceeding 124 kPa, as well as solvent oils, kerosene, diesel fuel, distillate fuel oils, lubricating oils, and other petroleum products.
Contact :
Product Details
I. Introduction
Dear customer, thank you very much for choosing the products manufactured by Jingzhou Chuangbo Instrument Manufacturing Co., Ltd. We are committed to providing you with the best possible service.
II. Overview
The copper strip corrosion tester is designed and manufactured in accordance with the People’s Republic of China’s petrochemical standard GB/T 5096, “Test Method for Copper Strip Corrosion of Petroleum Products,” and is suitable for determining the degree of copper corrosion caused by aviation gasoline, jet fuel, motor gasoline, natural gasoline, or other hydrocarbons with a Reid vapor pressure not exceeding 124 kPa, as well as solvent oils, kerosene, diesel fuel, distillate fuel oils, lubricating oils, and other petroleum products.
III. Main Technical Specifications
1. Power supply: AC 220 V ± 10%; 50 Hz
2. Temperature control range: Room temperature to 100.0°C
3. Temperature control accuracy: ±0.5℃
4. Temperature sensor: Pt100
5. Heating power: 800 W
6. Timing range: 0.01 seconds to 99 hours and 99 minutes
7. Sample-holding holes: 4
IV. Instrument Instructions
1. Place the instrument on the workbench, connect the interface cables, and install the sensor and thermometer. Ensure all switches are in the OFF position. Connect the power cord, making sure a reliable grounding wire is in place.
2. Add an appropriate amount of coolant to the tank.
3. Turn on the power switch and the temperature control switch; once the temperature controller displays normally, the system is ready to operate.
4. The upper row of digits on the temperature controller displays the bath’s actual measured temperature, while the lower row shows the setpoint temperature. Press the SET key and use the arrow keys < and the increase/decrease keys ∧∨ to adjust the temperature to your desired setting.
For operating the temperature controller, please refer to the attached Temperature Controller Manual.
Once the timer switch is turned on, the timer can be set to the desired test duration in hours (H) and minutes (M). When the preset time elapses, the buzzer will sound automatically to indicate that the time has expired. At this point, turn off the timer switch to stop both the display and the buzzer.
5. Different specimens require different test procedures:
A. Lubricating oil, solvent oil, and kerosene: Measure 30 mL of a sample that is completely clear, with no suspended water or entrained water, and transfer it into a clean, dry test tube equipped with a test tube holder. Place the final‑polished, clean copper strip into the sample in the test tube, then position the test tube—still held by its clamp—in a bath maintained at 100 ± 1°C. After keeping the test tube in the water bath for 3 hours ± 5 minutes, remove it and examine the copper strip.
B. Aviation gasoline and jet fuel: Measure 30 mL of a completely clear sample, free of suspended water or entrained water, and transfer it into a clean, dry test tube. Carefully slide the test tube into the test bomb, tighten the bomb cap, and place the bomb entirely in a bath maintained at 100 ± 1°C. After maintaining the bomb in the bath for 2 hours ± 5 minutes, remove it, rinse it with tap water for several minutes, open the bomb, extract the test tube, and inspect the copper strip.
C. Natural gasoline: Measure 30 mL of a completely clear sample, free of suspended water or entrained water, and transfer it into a clean, dry test tube. Carefully slide the test tube into the test bomb, tighten the bomb cap, and place the bomb entirely in a bath maintained at 40 ± 1°C. After keeping it in the bath for 2 hours ± 5 minutes, remove the test bomb, rinse it with tap water for several minutes, open the bomb, extract the test tube, and inspect the copper strip.
D. Diesel, fuel oil, and motor gasoline: Measure 30 mL of a completely clear sample, free of suspended water or entrained water, and transfer it into a clean, dry test tube equipped with a test‑tube holder. Place the final‑polished, clean copper strip into the sample in the test tube, then position the test tube—still held by its clamp—in a bath maintained at 50 ± 1°C. After keeping the test tube in the bath for 3 hours ± 5 minutes, remove it and examine the copper strip.
6. Inspection of Copper Sheets
1. Use tweezers to remove the copper sheet; rinse the sample with a cleaning solvent, then blot the solvent from the surface of the copper sheet using quantitative filter paper. Avoid touching the copper sheet with your hands. Placing the copper sheet in a flat observation tube will help prevent clamping marks and contamination during inspection and comparison.
2. Compare the copper specimen with the corrosion‑standard color chart, holding both at a 45° angle to the light so that refraction occurs, and observe carefully.
3. The corrosivity of the specimen is determined based on the grading of the standard corrosion color chart.
Grading of Corrosion Standard Color Charts
|
Grading |
Name |
Explanation |
|
1 |
Mild discoloration |
a pale orange, almost like freshly polished copper |
|
2 |
Moderate discoloration |
a purplish red |
|
3 |
Deep discoloration |
a multicolored hue with magenta overlaying brass tones |
|
4 |
Corrosion |
a transparent black, dark gray, or brown with only a hint of peacock green |
Explanation for determining corrosivity:
1. When the copper specimen falls between two adjacent standard color plates in terms of corrosion grade, the specimen shall be judged according to the more severely discolored corrosion grade.
2. When the copper sheet exhibits an orange hue darker than that of standard color chart shade 1b, it is still classified as Grade 1; however, if a reddish tint is observed, the copper sheet is rated as Grade 2.
3. Level 2’s medium purple copper sheet may be mistaken for Level 3, which appears brass-colored and is entirely covered by magenta. To distinguish between these two levels, immerse the copper sheet in a cleaning solvent: Level 2 will develop a deep orange color, while Level 3 will remain unchanged.
4. If the copper sheet becomes contaminated during heating and extraction, the experiment must be repeated.
5. If a corrosion level higher than that observed on most of the copper sheet’s surface develops along the edge or corner of the sheet’s plane, the test must be repeated.
6. If the two repeated measurements differ, the test shall be repeated. If the two results of the repeated test still differ, the specimen shall be classified according to the corrosion grade corresponding to the more severe discoloration.
V. Common Faults and Troubleshooting Methods
|
Fault Description |
Causes and Troubleshooting |
|
The indicator light is not on. |
The power plug is loose; the fuse has blown. |
|
The heater is on, but temperature control has failed. |
Solid-state relay failure |
|
Temperature display abnormal |
Temperature controller malfunction |
|
Does not heat up |
The thermostat or heating element is faulty. |