• 说明书-2.jpg

JCB0090-02 Engine Coolant Freezing Point Tester


This instrument is a low-temperature, precision testing device. Before use, please read the instruction manual carefully, operate with caution, and observe safety precautions.


Product Details


   I. Purpose and Scope of Application

  This instrument is designed and manufactured in accordance with the requirements specified in the People’s Republic of China standard GB/T 2430, “Method for Determining the Pour Point of Jet Fuel,” and the People’s Republic of China petrochemical industry standard SH/T 0090, “Method for Determining the Freezing Point of Engine Coolant.” By selecting different configurations, it can be used to determine the freezing point of jet fuel, engine coolant, and its concentrated solutions, making it a versatile freezing-point tester.

   II. Main Technical Specifications and Parameters

  1. Power supply: AC 220 V ±10%; 50 Hz.

  2. Working cold bath: Dewar flask, allowing continuous observation of the experimental process.

  3. Cold tank temperature control: +20℃ to -65℃.

  4. Temperature control accuracy: ±0.5℃.

  5. Bath solution stirring: The stirring motor operates automatically.

  6. Refrigeration System: Imported compressor cascade refrigeration.

  7. Sample stirring: electromagnetic stirring, (60–80) strokes per minute.

  8. Ambient temperature: ≤30℃.

  9. Relative humidity: ≤85%.

  10. Total power consumption: 1800 W.

   III. Main Structure

  This instrument consists of two parts: the right side houses the control and worktable, while the left side contains the compressor‑based refrigeration unit.

  Right side: Mainly includes the observable working Dewar flask, the stirring assembly, and the control panel.

  Left: Refrigeration workshop, housing refrigeration compressors and various components.

   IV. Instrument Installation

  1. Place the instrument on a level, stable workbench.

  2. Open the packaging box and verify the spare parts against the complete checklist provided in the instruction manual.

  3. It is recommended to use anhydrous alcohol as the refrigerant in the cooling tank.

  4. Plug the power cord into the rear panel of the instrument; the power supply must have a reliable grounding connection.

   V. Instructions for Use

   (1) Pre-test Preparation

  1. Before using this instrument, please read the instruction manual carefully.

  2. Before conducting the test, please carefully read the People’s Republic of China petrochemical industry standard SH/T 0090, “Method for Determining the Freezing Point of Engine Coolant,” and thoroughly understand and become familiar with the test method, procedures, and requirements specified in the standard.

  3. In accordance with the requirements specified in the standard, prepare all necessary test equipment, materials, and other items, and complete all preparatory work for the test.

  4. Verify that the instrument is operating under the environmental and operational conditions specified in the instruction manual.

  5. Inspect the instrument’s enclosure to ensure it is properly grounded.

   (II) Usage Instructions

   A. Preparation of Test Conditions

  1. Add an appropriate amount of alcohol to the Dewar flask.

  2. Insert the ice-point–specific test tube into the top test port of the Dewar flask and connect the experimental stirrer.

  3. Plug in the external power supply and turn on the power switch.

  4. Set the temperature control instrument to the desired temperature (e.g., −60°C) as follows: Press the SET key on the controller; the SV display will flash, allowing you to enter the required test temperature. Then press the SET key again to confirm the set temperature.

  5. Turn on the cooling switch to activate the instrument’s cooling mode. Once the temperature reaches the set control threshold, sample testing can proceed.

  Special Note:

  1. Before the instrument begins formal cooling, gently apply a small amount of glycerin to the exterior of the double-layer observation window using a clean cloth to prevent or reduce frost formation on the outside of the window.

  2. If frost or ice forms on the outer pane of the double‑glazed observation window and obstructs viewing, gently wipe it away with a clean cloth dampened with glycerin or alcohol.

  B. Freezing-point test for engine coolant and its concentrate

  1. Measure 85 mL of the sample and transfer it into a clean, dry test tube specifically designed for freezing-point determination. Simultaneously place the stirrer (a stainless-steel multi-loop type) into the test tube, secure the stopper, and insert the thermometer used for the test.

  2. Place the freezing-point tube, which has been loaded with the sample, stirrer, stopper, and thermometer, into the Dewar flask, then connect the stirrer to the upper stirring rod.

  3. Turn on the power switch, the cooling switch, the stirring switch, and the two green stirring buttons on the right, one after another.

  4. When the sample reaches the expected freezing point, record the temperature every 15 seconds.

  5. Plot the time–temperature curve on graph paper.

  6. On the time–temperature curve, identify the steeper portion (the cooling curve) and the flatter portion (the crystallization curve). The projection of the intersection point between the cooling curve and the crystallization curve onto the vertical axis corresponds to the freezing point of the sample (see Figure 3). If supercooling occurs, the highest temperature reached as the sample’s temperature rapidly rises after the onset of supercooling is taken as the freezing point of the sample (see Figure 4).

  7. During the test, the procedures specified in SH/T 0090-91 shall be strictly followed.

1

  Figure 3 Figure 4

   VI. Precautions

   1. Before plugging in the instrument, be sure to add an appropriate amount of anhydrous ethanol to the cooling bath; otherwise, do not power on the device under any circumstances!

  2. Since anhydrous alcohol undergoes thermal expansion and contraction at low temperatures, the amount added should be such that, once the apparatus has reached the set temperature and the test tube has been placed inside, the alcohol level does not exceed 2 cm below the top of the Dewar flask—please remember this! Be sure to replenish the cold bath with anhydrous alcohol regularly to compensate for evaporation losses, and promptly remove the test tube and the magnetic stirrer after the experiment is completed.

  3. If, after completing a low-temperature test, you wish to reset the bath temperature (to a value higher than the current bath temperature), you must first turn off the refrigeration switch. The switch may be turned on again only when the bath temperature has stabilized within approximately ±1°C of the setpoint.

  4. Regularly replace the alcohol in the Dewar flask to maintain clarity within the cold bath.

  5. To ensure proper operation of the compressor, after each startup of the compressor unit, unless an emergency arises, the refrigeration switch may be turned off only after 10 minutes; and after the compressor has been shut down, the refrigeration switch may be restarted only after another 10 minutes.

   Warning: If the instrument malfunctions, immediately disconnect the power supply. After the instrument’s cooling chamber has returned to room temperature, have it inspected and repaired by qualified personnel before resuming use to prevent accidents.

   VII. Instrument Complete Sets and Technical Documentation

  1. Complete instrument set

Serial number

Instrument Name

Unit

Quantity

Note

1

JCB0090-02 Freezing Point Tester

Q

1

 

2

Thermometer

-37 to 2°C, with a scale division of 0.2°C

support

1

 

-54 to -15°C, with a scale division of 0.2°C

support

1

 

-80 to 20°C, with a scale division of 0.5°C

support

1

 

3

Test tube (double-walled)

item

2

 

4

Spiral agitator

root

2

 

5

Test tube rack

set

2

 

6

Agitator connecting rod

root

2

 

7

Cross fastener

one

2

 

8

Test tube stopper (No. 6)

only

2

 

 

  2. Technical Documents

  ⑴ 1 copy of the instruction manual

  ⑵ One copy of the temperature controller user manual

  ⑶ One certificate of conformity


Production workshop


Aluminum recycling

Aluminum recycling

Aluminum recycling

Aluminum recycling

Aluminum recycling

Application Cases


Aluminum recycling

Aluminum recycling

Aluminum recycling

Aluminum recycling

Certificate of Honor


高新技术企业

高新技术企业

Core Supplier Certificate

Core Supplier Certificate

Occupational Health and Safety Management System Certification

Occupational Health and Safety Management System Certification

Environmental Management System Certificate

Environmental Management System Certificate

Quality Management System Certification

Quality Management System Certification

Tianjin Dongfang Credit Rating Certificate

Tianjin Dongfang Credit Rating Certificate

Related products


Leave a message


Welcome to leave an online message for inquiries and please keep your contact information up to date. We will get in touch with you as soon as possible.

%{tishi_zhanwei}%